Microwave Chemical Reactors Are Ideally Scientific And Practical

- Aug 05, 2019-

Microwave chemical reactors are ideally scientific and practical

Microwave chemical reactor adopts microcomputer technology and unique microwave adjustment technology to realize continuous adjustment of microwave power. It has a power meter display, which is easy to operate and has good repeatability. It is necessary for experiments that need to explore microwave catalytic reaction conditions, especially for those requirements. The chemistry of small microwave power and the desire to maintain continuous microwave radiation is ideally suited for its scientific and practical applications.

Microwave chemical reactor operation steps:

1. Turn on the microwave chemical reactor power and turn on the power switch. The power indicator light, the oven light, and the LCD screen light up at the same time.

2. The chemical that you need to react must be placed in the microwave cavity.

3. Set parameters for working hours

The reactor is divided into five working hours. The lower gear position, temperature and time of each section are independently adjustable.

The gear input range is 01-10, 01 is the low level of zui, indicating 10% power. 10 is zui, expressed as 100% power.

The temperature input range is 0-250, which means 0 °C to 250 °C.

The time input range is 0-999, which means 0 seconds to 999 seconds.

A cursor jumps on the screen to indicate the current input position.

Pressing the buttons on the touch panel automatically sounds “,”, indicating that the button is valid.

Press the 0-9 number keys on the touch panel to enter the corresponding character at the cursor position. If there is already a character at the cursor position, it will be overwritten by the character entered this time. At the same time, the cursor automatically moves back to the next input position.

Press the back panel of the touch panel to delete the character at the cursor position, and the cursor automatically advances to the previous input position.

Press the left and right cursor movement keys on the touch panel to move the cursor to the left and right to the input position.

Press the touch panel up and down cursor movement button, the cursor moves up and down in the vertical direction to the input position.

4. Running a microwave chemical reactor

Press the touch panel confirmation button to confirm that the set working period parameter is valid data. The LCD cursor will disappear and the new input data will no longer be accepted. The parameters confirmed this time will be saved in the flash memory and will not be lost after power off.

Press the touch panel open button, the microwave chemical reactor starts working according to the user setting, and the microwave starts to launch. The microwave output is displayed in the lower right corner of the LCD screen. The LCD displays the current actual temperature, effective power, and actual time in real time.

Press the touch panel close button, the microwave chemical reactor stops working, and the microwave is no longer emitted. The microwave in the lower right corner of the LCD screen has stopped.

Modify the reaction conditions during the operation of the microwave chemical reactor: press the touch panel setting button, the LCD cursor appears again, you can modify the parameters. After the modification, you must press the touch panel confirmation button to confirm. You don't have to press the Open button again.

5. Magnetic stirrer control

The control of the magnetic stirrer is controlled by the microcomputer alone, and the knob on the right side of the microwave chemical reactor door is rotated clockwise, and the speed of the magnetic stirrer is increased, and vice versa.